Application
The CLS54D conductivity sensor is specifically designed for use in hygienic
applications in the food, beverage and pharmaceutical industries and in
biotechnology. Thanks to its hygiene certificates and its food-safe, virgin PEEK
design without any joints or crevices, it meets the strict demands of these industries.
The CLS54 is ideal for:
- Phase separation of product/water and product/product mixtures in pipe systems
- Control of cleaning in place (CIP) processes in the return line
- Concentration control in the remaking of CIP cleaning agents
- Product monitoring in pipes, bottling plants and quality assurance
- Leakage monitoring
in the following industries:
- Dairies
- Breweries
- Beverages (water, juices, soft drinks)
- Pharmaceuticals and biotechnology
Use with Liquiline CM42, CM44x and CM14 transmitters.
Your benefits
- Unique hygienic design, therefore no risk of recontamination
- Has all the process connections commonly used in the hygiene sector
- Fast measurement with temperature response time t 90 under 26 s, ensuring safe and efficient phase separation
Other advantages of Memosens technology
- Maximum process safety
- Data security thanks to digital data transmission
- Very easy to use as sensor data saved in the sensor
- Recording of sensor load data in the sensor enables predictive maintenance
Measuring principle
An oscillator (1) generates an alternating magnetic field in the primary coil (5), which induces a
current flow (4) in the medium. The strength of the current depends on the conductivity and thus on
the ion concentration in the medium. The current flow in the medium, in turn, generates a magnetic
field in the secondary coil (3). The resulting induced current is measured by the receiver (2) and used
to determine the conductivity.