Application
- Chemical processes
- Pulp and paper industry
- Contaminated media:
- Solids
- Emulsions
- Precipitation reactions
- Dispersions
Your benefits
- Open aperture for use in contaminated media
- Low maintenance thanks to gel filling
- Long service life thanks to new, stabilized gel
- Not affected by fluctuations in pressure and temperature
- Fast response time
Other advantages of Memosens technology
- Maximum process safety thanks to non-contact, inductive signal transmission
- Data security thanks to digital data transmission
- Very easy to use as sensor data saved in the sensor
- Recording of sensor load data in the sensor enables predictive maintenance with the Memobase Plus CYZ71D
Measuring principle
ORP measurement
The ORP potential is a unit of measurement for the state of equilibria between oxidizing and
reducing components of a medium. The ORP is measured using a platinum or gold electrode instead
of the pH-sensitive glass membrane. Analog to the pH measurement, an integrated Ag/AgCl
reference system is used as a reference electrode.
Measuring system
A complete measuring system consists of the following components at least:
- ORP electrode CPS92D or CPS92
- Transmitter, e.g. Liquiline CM42, CM44x/R, Mycom S CPM153, Liquisys M CPM2x3
- Memosens data cable CYK10 for Memosens sensors or CPK9 for analog sensors
- Immersion, flow or retractable assembly, e.g. Cleanfit CPA871/875