Automatic sample conditioning system for supplying
process measuring devices with filtered sample from the
inlet or sludge activation phase of a wastewater
treatment plant
Application
Liquiline System CAT860 enables the fully automatic
collection and filtration of aqueous samples. Thanks to the
modular concept, the sample preparation system can be
adapted to various different process conditions. There are
various order options available for this purpose. For installing
the filter in the process medium, various assemblies are
available as accessories. The heated version of the hoses and of
the housing mean that the system can be used at temperatures
as low as -20 C.
The device is designed for use in the following applications:
- Collection of aqueous samples from industrial processes
- Communal and industrial wastewater treatment plants
Your benefits
- Robust and reliable:
- Peristaltic pump, sampling takes place free from interference
- Samples free from particles and algae thanks to membrane filtration
- Automatic filter backflushing with cleaner
- Filter backflushing with air
- Optional hose trace heating system
- Simple and user-friendly:
- Control via a CA80 analyzer
- Installation compatible with the Flexdip CYH112
- Cleaning and maintenance: Replacement of pump hoses and filter without using tools
Function
A peristaltic pump is used to take the sample. The sample is drawn in via the filter element, filtered
and fed to the analyzer. The automatic cleaning function backflushes the filter and the sample hoses
with a cleaning solution. Backflushing with a cleaning solution is performed using a vacuum pump.
The cleaning valve is switched for this purpose. It is also possible to backflush the filter and the
sample hoses with compressed air.
Measuring system
A complete sample conditioning system comprises:
- Liquiline System CAT860 sample preparation system
- Local operation with soft keys and status LEDs
- Filter unit with filter and assembly in the configuration ordered
- Automatic cleaning function with compressed air (external compressed air supply necessary)
- Peristaltic pump for pumping the sample
- Vacuum pump for the automatic chemical cleaning function
- Housing heating (optional)
- Sample hose, filter to pump in the configuration ordered, optionally heated
- Sample hose, pump to analyzer in the configuration ordered, optionally heated
- Cleaner (to be ordered separately)